DSpace Repository

On Accuracy of Radiotherapy Dose Delivery

Show simple item record

dc.contributor.author Eman Abdelrazig Mohamed Osman
dc.date.accessioned 2018-12-17T06:33:23Z
dc.date.available 2018-12-17T06:33:23Z
dc.date.issued 2017
dc.identifier.citation International University of Africa - Dean of Graduate Studies & Scientific Research & Publication - Faculty of Pharmacy en_US
dc.identifier.uri http://dspace.iua.edu.sd/handle/123456789/3930
dc.description.abstract This study simulates possible errors – in radiotherapy- which could be happened, to determine their effect on delivereddose.And to know which of them have most effect.Nine experiments were done to estimate the error in radiotherapy, four of them in two depth (5,10)cm and another in one depth (10)cm Cobalt60 machine, ionization chamber, electrometer and slab phantom were used for all nine experiments. It has been founded that the most effective parameter was lateral shift and the least one was gantry shift.Lateral and longitudinal shifts had the same effect in depth (5)cm.Then SSD shift followed by FS shift.For depth (10)cm lateral shift followed by temperature shift.For the FS,SSD, and longitudinal shifts they had the same effect at first values.Then came thecollimator shift and finally thegantry shift. en_US
dc.language.iso en en_US
dc.subject medical physics en_US
dc.title On Accuracy of Radiotherapy Dose Delivery en_US
dc.type Thesis en_US


Files in this item

This item appears in the following Collection(s)

Show simple item record

Search DSpace


Advanced Search

Browse

My Account